INSTALLATION REQUIREMENTS FOR THE
ULTRA-HIGH PRECISION STEREOCOMPARATOR
I. INSTRUMENT
A. Name of instrument:
B. Manufacturer:
C. Contract number:
D. Delivery date:
Ultra-High Precision Stereocomparator
Tentative: January 12, 1970
II. PHYSICAL FEATURES
A. Sub-assemblies:
1. Number of sub-assemblies: NA
Main frame
2. Largest sub-assembly: Weight 6000 lbs; 18" H x 123" W x 74" D
3. Heaviest sub-assembly: Weight 6000 lbs; 18" H x 123" W x 74" D
B. Assembled instrument:
1. Number of major components: 7
2. Largest component: Weight 27000 lbs; 9
3. Heaviest component: Weight 27000 lbs; 9
4. Total floor space required after assembly,
space: 10' H x 18' W x 28' D
S. Total weight of assembled instrument: 32,
C. Type of base of mount: 4-point suspension
D. Does the instrument have built-in mobility? No.
E. Is the instrument particularly sensitive to vibration? Yes.
Will the instrument generate vibration? No.
F. Are any special or unusual tools or fixtures necessary or advisable for the
installation or the maintenance of this instrument? No (mechanical).
III. UTILITIES
A. Electrical:
1. Voltage
2. Current
3. Frequency
4. Nr. of phases
5. Nr. of wires
6. Power Required
7. Power factor
8. Type of outlet:
9. Type of ground:
10.
0" H x 160" W x 96" D
0" H x 160" W x 96" D
including maintenance access
000 lbs.
110/120 volts A.C., 208 V A.C., 24 V D.C.
100 Amps/phase
60 cps
3 Ph
4
40,000 watts
.7 (Lagging)
Permanent v 01 +
Direct earth ground.
Should the instrument be shielded, either from external electromagnetic
signals or to prevent interference with other equipment? Yes.
If "Yes," to what extent? Machine will be self-shielded.
STAT
44-46 pke,
-1-
Declassified in Part - Sanitized Copy Approved for Release 2012/08/29: CIA-RDP79B00873A001300010004-6
Declassified in Part - Sanitized Copy Approved for Release 2012/08/29: CIA-RDP79B00873A001300010004-6 FAT
?
-4-May 1968
14.7 FT/ lto?
1 IT &
Air conditioning:
1. Desired environment: Room air temperature of 72?F ? 0.50F and
relative humidity of 66% ? ST: , (NA- )
2. Input Air: Is a direct connection necessary? Yes.
Required input air temperature 72?F ? 0.5?F.
Relative humidity 55% J.:
Maximum particle size in entering air less than 1 micrometer diameter.
3. (a) Film cooling air - Intermediate pressure
35 psig - 4 cfm, 20 watts to dissipate, total,
To match room ambient air.
(b) Lamp cooling air - Low pressure
5" H2iD at supply, 2-1/2" H20 at distribution header, 1/2" change
in pressure throughrequipment STAT
(
I) 450 watts = 130 cfm for 2 lamps with 500F rise)
2) 75 watts = 40 cfm (for 2 lamps with 50?F rise)
To match room ambient air
3) Optical bridge
Total 4 watts to be dissipated at 1/4?F rise
To match room ambient air
100 cfm total
1" H20 at header - No return system
(c)
Cabinet cooling
1) (Electronic) - #1
- 2000 cfm
2) (Electronic) - 42
- 1300 cfm
2-1/2" H20 at distribution header
3) (Electronic) - 413
- 2400 cfm
f
1/2" change in pressure through
4) (Utilities) - #4
- 450 cfm
NRI equipment
(d)
Sub-Platen Cooling
25 watt total, 1/4?F rise - 300 cfm (2 platens), 1" H20 at header - no
return. To match room ambient.air
rf) Teletype - 5 amps at 110 volts = 550 watts
) Card Punch - 5 amps at 110 volts = 550 watts
(g) Control console - 275 watts
10F rise - 1000 cfm - Pressure change = 1/2" H20 through NRI equipment.
Supplied with room ambient air
(h) People
2 operators, 2 visitors, 2 engineers, 2 service men, 4 supervisors.
Total of 12 people.
(i) Air-Bearing
1 cfm - Refrigerate to room ambient air - 150 psig COMPRESSOISTAT
No return system
(j) Film lift off
35 cfm - Room ambient air - 35 psig - No return system
(k) Vacuum system - PUMP) STAT
(1) Room ambient conditions
72?F? 0.5?F - 60% RH -10% RH - Particle count, class 100 clean room
NOTE: Atmospheric pressure at
C. No water is required.
equipment.
D. Compressed air:
Is compressed air required? Yes. Water free and oil free.
Pressure 35 psig.
Flow in CFM maximum IQ, minimum 1, average 4.
-2-
STAT
Declassified in Part - Sanitized Copy Approved for Release 2012/08/29: CIA-RDP79B00873A001300010004-6
Declassified in Part - Sanitized Copy Approved for Release 2012/08/29: CIA-RDP79B00873A001300010004-6 STAT
14 Ivi.ity 1968.
E., No vacuum is required.
F. Peripheral Devices:
The instrument is connected to a peripheral computer by a cable and connector
mating wit cannon connector number RWK-8-31SL. Connections are STAT
specified by logic diagram, output interface, number EE 5832. STAT
-3-